A 4096 channel event-based multielectrode array with asynchronous outputs compatible with neuromorphic processors


Creative Commons License

Cartiglia M., Costa F., Narayanan S., Bui C. H., Ulusan H., Risi N., ...Daha Fazla

NATURE COMMUNICATIONS, cilt.15, sa.1, 2024 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 15 Sayı: 1
  • Basım Tarihi: 2024
  • Doi Numarası: 10.1038/s41467-024-50783-2
  • Dergi Adı: NATURE COMMUNICATIONS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, BIOSIS, CAB Abstracts, Chemical Abstracts Core, Geobase, INSPEC, MEDLINE, Veterinary Science Database, Directory of Open Access Journals
  • Orta Doğu Teknik Üniversitesi Adresli: Evet

Özet

Bio-signal sensing is pivotal in medical bioelectronics. Traditional methods focus on high sampling rates, leading to large amounts of irrelevant data and high energy consumption. We introduce a self-clocked microelectrode array (MEA) that digitizes bio-signals at the pixel level by encoding changes as asynchronous digital address-events only when they exceed a threshold, significantly reducing off-chip data transmission. This novel MEA comprises a 64 x 64 electrode array, an asynchronous 2D-arbiter, and an Address-Event Representation (AER) communication block. Each pixel operates autonomously, monitoring voltage fluctuations from cellular activity and producing digital pulses for significant changes. Positive and negative signal changes are encoded as "up" and "down" events and are routed off-chip via the asynchronous arbiter. We present results from chip characterization and experimental measurements using electrogenic cells. Additionally, we interface the MEA to a mixed-signal neuromorphic processor, demonstrating a prototype for end-to-end event-based bio-signal sensing and processing.