An interleaved drive electrical impedance tomography image reconstruction algorithm


EYÜBOĞLU B. M.

PHYSIOLOGICAL MEASUREMENT, cilt.17, 1996 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 17
  • Basım Tarihi: 1996
  • Doi Numarası: 10.1088/0967-3334/17/4a/009
  • Dergi Adı: PHYSIOLOGICAL MEASUREMENT
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Orta Doğu Teknik Üniversitesi Adresli: Hayır

Özet

In this study, a reconstruction algorithm for a 16-electrode interleaved-drive electrical impedance tomography (EIT) system is developed, based on inversion of an analytically calculated sensitivity matrix. The sensitivity matrix is calculated using Geselowitz's lead-sensitivity theorem. Eight interleaved electrodes out of 16 (equally spaced) electrodes are designated as current injection electrodes and the remaining eight electrodes are designated as measurement electrodes. The sensitivity matrix is singular, therefore singular value decomposition (SVD) of the sensitivity matrix, followed by pseudoinversion-with and without truncation-is used to reconstruct images. The algorithm is a single-pass algorithm. Data from a saline filled tank and in viva data during respiration and the cardiac cycle, acquired by using a Sheffield multifrequency system, are used to reconstruct images. The effect of different truncation levels on the reconstructed images is investigated.