Stibine preconcentration in a quartz trap with subsequent atomization in the quartz multiatomizer for atomic absorption spectrometry


Korkmaz D., Dedina J., Ataman O.

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, cilt.19, sa.2, ss.255-259, 2004 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 19 Sayı: 2
  • Basım Tarihi: 2004
  • Doi Numarası: 10.1039/b308883a
  • Dergi Adı: JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.255-259
  • Orta Doğu Teknik Üniversitesi Adresli: Hayır

Özet

A preliminary evaluation of a simple (bare) quartz tube trap for collection of SbH3 and for volatilization of trapped analyte with subsequent atomization in a multiple microflame quartz tube atomizer (multiatomizer) for atomic absorption spectrometry is presented. The influence of relevant experimental parameters on the collection/volatilization efficiency was investigated. The parameters studied Were: collection temperature and time, volatilization temperature and flow rates of air and hydrogen for the volatilization. Under optimized conditions, the collection/volatilization efficiency was 65%. For the collection time of 120 s (sample volume of 8 ml), the detection limit was 3.9 pg ml(-1), and the analytical plot was linear up to 1.25 ng ml(-1). Possible ways of improving the performance of the set-up as well as the collection/volatilization efficiency are suggested.