Single Current Diffusion Tensor Magnetic Resonance Electrical Impedance Tomography: A Simulation Study
ISMRM & SMRT Virtual Conference & Exhibition, Sydney, Australia, 8 - 14 August 2020, vol.1, no.1, pp.1-3, (Full Text)
- Publication Type: Conference Paper / Full Text
- Volume: 1
- City: Sydney
- Country: Australia
- Page Numbers: pp.1-3
- Middle East Technical University Affiliated: Yes
Abstract
Diffusion Tensor Magnetic Resonance Electrical Impedance Tomography (DT-MREIT) is an imaging technique providing low-frequency conductivity tensor images. In all DT-MREIT applications in the literature two linearly independent current injection patterns are used to reconstruct the extra-cellular conductivity and diffusivity ratio (
) which is the space-dependent scale factor between
and
in a porous medium. In this study, a new approach is proposed to reconstruct
using only one current injection pattern. The proposed method is evaluated using a simulated measurements with additive noise. The obtained results show similar performance compared to the conventional two current DT-MREIT.