An experimental analysis of the stresses on the implant in an implant-tooth-supported prosthesis: A technical note


Uysal H., Iplikcioglu H., Avci M., Bilir O., Kural O.

INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, cilt.12, sa.1, ss.118-124, 1997 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 12 Sayı: 1
  • Basım Tarihi: 1997
  • Dergi Adı: INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.118-124
  • Orta Doğu Teknik Üniversitesi Adresli: Hayır

Özet

When a natural tooth connected to an implant abutment by a framework is displaced within physiologic limits under functional loads, the superstructure acts as a cantilever moment arm and the implant is loaded additionally during that moment effect. The effects of a deflecting support connected to art IMZ implant, with either the IMC resilient or the titanium rigid element, were evaluated through an experimental model. The strain values measured when the moment arm was not allowed to deflect were only 30% lower than the strains measured when it was deflected within the limits of physiologic tooth mobility. The IMC was effective and achieved a reduction in the cervical strains up to 60% when compared with the rigid connector.